13.0 Write Calibration to Flash

Equipment Needed
Power Supply 12V 10A,
Battery Current Test Fixture (P/N 9100-0690-TF) or AED PRO Disposable Battery (P/N
1008-002003-01),
USB Flash Drive (P/N 8000-000474),
AED PLUS, Quick-Test, Test Fixture (P/N 9100-0567-TF).
Setup:
Connect AED PLUS Quick-Test Test Fixture (P/N 9100-0567-TF) to the unit.
Place the in the battery well.
Set the Power Supply to 12.00V.
Turn on the Power Supply and turn the unit off.
Note: To download data to a USB device in the AED pro, make sure you are in
Test
13.1
13.2
13.3
13.4
1–20
T
ESTS
Non-Rescue Mode.
Action
Press and hold the On/Off
button for at least 7 seconds,
then release it.
Attach a USB device to the port
inside the battery compartment,
and then press the USB softkey.
Press the ALL softkey to
download all patient data on the
unit.
www.zoll.com
Expected Result(s)
Verify the unit issues a voice
and text prompt UNIT OK,
Verify the unit issues a voice
and text prompt USB DEVICE
CONNECTED, followed by
DOWNLOADING DATA and
DATA DOWNLOAD
COMPLETE
If there is no USB device
attached or if the unit fails to
establish a connection with the
device, the unit issues command
INSERT USB DEVICE.
Pass
Fail
9650-0309-01 Rev. E