14.0 Audio Recording Test (if option installed)

Equipment Needed
Power Supply 12V 10A,
AED PLUS, Quick-Test, Test Fixture (P/N 9100-0567-TF),
USB Flash Drive (Clinical Files from previous step),
PC with RescueNet Code Review.
Setup:
Connect AED PLUS Quick-Test Test Fixture (P/N 9100-0567-TF) to the unit.
Place the in the battery well.
Set the Power Supply to 12.00V.
Turn on the Power Supply and turn the unit off.
Note: The AED Pro is capable of recording up to 20 minutes of continuous
Test
14.1
14.2
14.3
9650-0309-01 Rev. E
ECG/Audio recording if Audio Recording is enabled during a rescue.
Action
Insert USB Flash Drive
containing clinical event data
(imported from the AED in USB
Test Step) in computer USB
Slot. Open file explorer, and
select USB Drive
Double-click on desired file and
then click Yes on the messaging
box
Select "3-Magnified ECG" from
the top of the screen and click on
the Play button
ZOLL AED Pro Service Manual
Expected Result(s)
Verify files are present
Verify a RecueNet Code Review
, Standard Edition screen pops
up
Verify that all voice prompts
and events from the device itself
recorded in the environment can
be heard from the PC speaker.
Test Equipment
Pass
Fail
N/A