9.0 Patient Impedance Calibration

Equipment Needed
Power Supply 12V 10A,
Battery Current Test Fixture (P/N 9100-0690-TF) or AED PRO Disposable Battery (P/N
1008-002003-01),
AED PLUS Quick-Test Test Fixture (P/N 9100-0567-TF),
PC with AED PLUS/PRO DCI Panel software (P/N 9103-0815-SW),
AED PRO Resistance, Decade Box (P/N 9100-000957-TF) or equivalent (e.g.RS-200
Resistance Substituter).
Setup:
Power up with IrDA Mode (communication mode).
Note: For best transmission results, the ZOLL USB IrDA port (P/N 9355-0807) must be
facing the IrDa port of the unit with a clear 10 to 18-inch line-of-sight.
Note: Patient Impedance and ECG Calibration are saved together. To test calibration of both
Impedance and ECG offset, select DCI commands: GICAL; GEVOL, 2000; GECAL.
Connect AED PLUS Quick-Test Test Fixture (P/N 9100-0567-TF) to the unit.
Connect the Decade Box to the HV Connector Cable of the AED PLUS Quick-Test Test
Fixture (P/N 9100-0567-TF) to set the appropriate resistance for Impedance calibration. The
decade box can only handle ¼ watt, DO NOT discharge into the box.
Test
9.1
9.2
9.3
9.4
9650-0309-01 Rev. E
Action
Double click on DCI_Panel icon
on Desktop. Select AED Pro
mode.
Set Decade Box to 10 Ohms.
Select SIL (hi, lo, md) from the
'Command 1' window, enter
LO in 'Parm 1 (val)' window
and click SEND.
Set Decade Box to 35 Ohms.
Select SIL (hi, lo, md) from the
'Command 1' window, enter
MD in 'Parm 1 (val)' window
and click SEND.
Set Decade Box to 300 Ohms.
Select SIL (hi, lo, md) from the
'Command 1' window, enter HI
in 'Parm 1 (val)' window and
click SEND.
ZOLL AED Pro Service Manual
Expected Result(s)
Verify the DCI Panel.vi appear
screen is displayed.
The response will be OK.
The response will be OK.
The response will be OK.
Test Equipment
Pass
Fail
1–13