Power Supply Test (Optional); Equipment Needed; Set Up - ZOLL AED 3 Service Manual

Hide thumbs Also See for AED 3:
Table of Contents

Advertisement

Power Supply Test (Optional)

This test is comprised of two tests: the Off Current test and the On Current test.
In order to perform the Off Current test, two measurements are required: one
with the fixture installed in the device and one for the test fixture. The difference
of the two readings is the Off Current Value.
NOTE If a Power Supply is not available for this test, record N/A for this test

Equipment Needed

Battery Test Fixture, Power Supply, Digital Multi-Meter.

Set Up

1. On the Digital Multi-Meter, set the function to DC Amps.
2. Connect the positive lead from the I jack to the positive current post (Amp
meter) on the Battery Test Fixture.
3. Connect the negative lead from the LO jack to the negative current post on
the Battery Test Fixture.
4. Set the Power Supply to 12.0V, then turn it off.
5. Connect the leads from the Battery Test Fixture to the proper polarity of the
Power Supply and install the Battery Test Fixture into the battery well.
Action
1.
Turn on the Power Supply.
2.
Verify that the AED is turned off to
perform the Off Current test.
On the Battery Test Fixture, turn
the I Measure switch to the sleep
mode position.
After the measurement is taken,
turn the I Measure switch back to
the off position.
3.
Remove the Battery Test Fixture
from the unit.
Turn the I Measure switch to the
sleep mode position.
After the measurement is taken,
turn the I Measure switch back to
the off position.
8
on the Maintenance checklist on the back page of this manual.
ZOLL AED 3 Service Manual
Expected Result(s)
Verify that the AED runs a self test
(approximately one minute) and
then powers off.
Note: AED may issue the messages
CHANGE BATTERY and UNIT FAIL.
Disregard the messages and
continue the test.
Write down the current
measurement on the Digital
Multi-Meter.
Value #1______________mA
Write down the current
measurement on the Digital
Multi-Meter.
Value #2______________mA
Pass/Fail
o o
9650-000761-01 Rev. A

Hide quick links:

Advertisement

Table of Contents
loading