ZOLL AED Pro Service Manual page 20

Hide thumbs Also See for AED Pro:
Table of Contents

Advertisement

C
1 M
HAPTER
AINTENANCE
5.7
5.8
5.9
5.10
5.11
5.12
5.13
5.14
.
1–8
T
ESTS
Press VFIB on the Defibrillator
Analyzer.
Press Flashing Shock button
nd
2
shock energy test (default
value 150 joules)
Press VFIB on the Defibrillator
Analyzer.
Press Flashing Shock button
rd
3
shock energy test (default
value 200 joules)
Press and hold the On/Off
button to power off the unit.
www.zoll.com
After 2 minutes (default setting),
Verify the unit prompts STOP
CPR, followed by DON'T
TOUCH PATIENT
ANALYZING twice, followed
by SHOCK ADVISED,
DON'T TOUCH PATIENT,
PRESS FLASHING SHOCK
BUTTON.
Verify that unit prompts
SHOCK DELIVERED.
Verify the Energy delivered into
analyzer is 122 joules to 166
joules.
Verify the unit prompts START
CPR, followed by CONTINUE
CPR.
After 2 minutes (default setting),
Verify the unit prompts STOP
CPR, followed by DON'T
TOUCH PATIENT
ANALYZING twice, followed
by SHOCK ADVISED,
DON'T TOUCH PATIENT,
PRESS FLASHING SHOCK
BUTTON.
Verify that unit prompts
SHOCK DELIVERED.
Verify the Energy delivered into
analyzer is 155 joules to 211
joules.
Verify unit displays UNIT
POWERING OFF message.
9650-0309-01 Rev. E