ZOLL AED Pro Service Manual page 26

Hide thumbs Also See for AED Pro:
Table of Contents

Advertisement

C
1 M
HAPTER
AINTENANCE
9.5
9.6
1–14
T
ESTS
Remove the Decade Box.
Select GICAL from 'Command
1' window and click SEND.
Select SFCAL from 'Command
1' window and click SEND.
www.zoll.com
The response will be SILLO,
SILMED, SILHI. Compare as
followed:
56 < SILLO< 176 counts (10
Ohms),
134 < SILMD< 321 counts (35
Ohms),
972 < SILHI< 2129 counts (300
Ohms).
The response will be SFCAL,
OK.
Note: the Calibration values will
go back to default conditions if
not saved. Save calibration
values. both ECG offset and
Impedance must be calibrated.
9650-0309-01 Rev. E