Table 2-4. AED Pro Error Log Messages (continued)
Error
No.
14
15
17
18
19
20
21
22
23
24
25
26
27
28
29
Note: Error numbers 256 and greater indicate noncritical conditions.
9650-0309-01 Rev. E
Error Message
Patient impedance experienced a short
The A2D is getting non-accurate readings during its test
The message queue overflowed
PowerOn task has timed out while waiting for a self test result to come back.
Calibration faults that occur during a clinical event.
An unexpected error has occurred when calling a Nucleus operation.
An unexpected error has occurred when calling a Rhapsody operation.
An unexpected coding error has occurred.
No Language installed - corrupt language.
The shock button is pressed/stuck.
The ECG Calibration data is invalid
The Patient Impedance Calibration data is invalid
A critical flash error has occurred
The WatchDog is not enabled during clinical event
The clock test failed
ZOLL AED Pro Service Manual
AED Pro Error Log Messages
2–7