C
1 M
HAPTER
AINTENANCE
12.0 Real Time Clock Test
Equipment Needed
•
Power Supply 12V 10A,
•
Battery Current Test Fixture (P/N 9100-0690-TF) or AED PRO Disposable Battery (P/N
1008-002003-01),
•
AED PLUS Quick-Test Test Fixture (P/N 9100-0567-TF).
Setup:
•
Connect AED PLUS Quick-Test Test Fixture (P/N 9100-0567-TF) to the unit.
•
Place the Battery Current Test Fixture (P/N 9100-0690-TF) in the battery well.
•
Set the Power Supply to 12.00V.
•
Turn on the Power Supply and turn the unit off.
Note: To set the current date and time in the AED pro, make sure you are in
Test
12.1
12.2
1–18
T
ESTS
Non-Rescue Mode.
Action
Press and hold the On/Off
button for at least 7 seconds,
then release it.
Press SET TIME softkey.
www.zoll.com
Expected Result(s)
Verify the unit issues a voice
and text prompt UNIT OK,
Verify the unit prompts you to
set the Year, Month, Day Hour,
Minute, and Second. To advance
to each subsequent screen, press
softkey.
Pass
Fail
9650-0309-01 Rev. E