The accuracy for measuring open circuit voltage is given as:
A
where
The value of A
The values of the other factors are given below.
where
During a VOCX test, the AT3600 automatically applies the test voltage to the part
under test and checks that the voltage has stabilised before recording measurements.
The time that the voltage takes to stabilise will vary with the source type used and
the nature of the part under test.
In general (and especially for source types 'Programmable' and 'AT Output
Transformer'), the AT3600 will record a measurement very quickly, (typically in less
than 0.5 seconds) which is desirable in a production test environment.
If the measurement were to be recorded after a longer period of time under power,
slightly different readings may be obtained due to self-heating effects in the part
under test. This effect may be noticed when comparing results taken using different
source types or with slower measurement systems.
S
PECIFICATION
(E
EST
XTERNAL
= A
+ A
+ A
+ A
C
V
T
R
A
is the total accuracy (%)
T
A
is the basic relative accuracy (%)
R
A
is the calibration accuracy (%)
C
A
is the correction for voltage level (%)
V
A
is the correction for the external source type (%)
SRCE
is given in the table in section 1.1 of this chapter
R
A
=
0.03% + (0.001% x
C
A
=
0.03% + (0.001% /
V
A
= 0.5% for manual / line supply
SRCE
= 0.0% for analogue / amplifier and AT output transformer
F
is the measurement frequency (Hz)
M
V
is the measured voltage (V)
M
S
)
OURCE
SRCE
F
)
M
V
)
M
VPN 98 - 024
S
PECIFICATION
P
8.1.11.