Voltech AT3600 User Manual page 216

Automatic transformer tester
Hide thumbs Also See for AT3600:

Advertisement

T
R
H
EST
ESULTS
ANDLING
P
4.3.26.
AGE
Float value giving the minimum limit for the test TestNumber.
"NO LIMIT"
GetMaxLimit (LPCSTR ComPort, short TestNumber)
Float value giving the maximum limit for the test TestNumber.
"NO LIMIT"
GetTestStatus (LPCSTR ComPort, short TestNumber)
Returns four digit hex string ("####") giving the test status as seen on editor
results screen.
GetTestMnemonic (LPCSTR ComPort, short TestNumber)
Returns the test mnemonic as a string, i.e. "LBAL".
GetTestUnits (LPCSTR ComPort, short TestNumber)
Returns the test units as a string, i.e. "Ohm".
GetCompensationStatus (LPCSTR ComPort)
"NO COMPENSATION STATUS"
"No Fixture Compensation"
"Fixture Compensation: successful"
"Fixture Compensation: fail" + ExtraString
GetUnitType(LPCSTR ComPort)
Returns "AT3600", "AT1600" or "ATi" depending on the type of instrument
connected to the ComPort.
GetResult (LPCSTR ComPort, short TestNumber)
The number returned is a float representing the value of the result measured or a
zero if no result was available.
This function is included for compatibility with older versions of the Voltech AT
Server. It is not recommended for new applications.
This test has no minimum limit.
This test has no maximum limit.
VPN 98 - 024
V
AT3600 U
OLTECH
SER
No status received yet.
Program has not been
compensated.
Compensation all OK.
Compensation has failed.
U
T
SING
HE
M
ANUAL
S
ERVER