Section 6

Calibration Preparation

Perform the following steps to prepare the DBV-
500 for Sensor Calibration:
1. Remove electrical power from the DBV-500
Unit.
2. Set DIP Switch1 #8 to ON (Figure 6-25).
3. Apply electrical power to the DBV-500 Unit. The
Status LED will begin flashing at a
rate.

Figure 6-25 DIP SW1 Setting

ON
1
2
3
Figure 6-25 DIP SW1 Setting
4. Launch the "
JCM Tool Suite Standard Edition
Application (Figure 6-26).
Figure 6-26 Launching JCM Tool Suite Stan-
dard Edition Screen
Figure 6-26 Launching JCM Tool Suite Standard
Edition Screen
5. Click the "
Service Mode
select "
Sensor Adjustment

Figure 6-27 Sensor Adjustment Selection

Figure 6-27 Sensor Adjustment Selection
6. Confirm that the Sensor Calibration Program
Screen appears (Figure 6-28).
Figure 6-28 Sensor Calibration Screen 1
Figure 6-28 Sensor Calibration Screen 1
P/N 960-100933R_Rev. 2 {EDP #233449}
DBV™ Series DBV-500 Banknote Acceptor
Green
Color
4
5
6
7
8
"
" Pull-Down Menu, and
" (Figure 6-27 a).
a

Sensor Calibration

To calibrate the DBV-500 sensors, proceed as 
follows:
1. Click the "
Sensor Calibration
6-29 a).
Figure 6-29 Sensor Calibration Screen 2
a
Figure 6-29 Sensor Calibration Screen 2
2. Confirm that the "
" Screen appears (Figure 6-30).
TION
3. Referring to the screen, confirm that the Trans-
port Path is locked firmly and click the "
button (Figure 6-30 a) to begin the non-paper
calibration.
NOTE: Make sure there are no
foreign objects in the Transport Path.

Figure 6-30 Non-Paper Calibration Screen

a
Figure 6-30 Non-Paper Calibration Screen
NOTE: When the Calibration does
not complete with an expected
Positioning Sensor D/A value, a
warning message window pop up will
appear. Click "YES" Button to
proceed.
© 2020 JAPAN CASH MACHINE CO., LTD.
6 - 6
Calibration and Testing
"
button (Figure
DBV-500 SENSOR CALIBRA-
"
Start