More About Referencing

Out of Range Condition

Manually Setting the Baseline at the Default Value

Manually Overriding the Baseline Default Value

Automatic Baseline "Zeroing"

6-4
Uterine Activity Monitoring: Tocotransducer (External Method)
After you press the
least 100 relative units above the reference level (probably because the belt is too
tight), the UA display area flashes "– –". If this happens, remove the tocotransducer
from the patient; re-reference with no pressure applied to the button; re-apply the
transducer to approximately 25 relative units above the baseline; then re-reference
one more time. If you still receive the same result, try a different tocotransducer or
contact your Service Representative.
Briefly pressing the
default is configured via the service setup mode. The monitor is shipped from the
factory with a default setting of 10 relative units. Qualified service personnel can
access the service setup mode to set the baseline default to 5, 10, 15, 20, or 25
relative units. Refer to the
Pressing and holding the
UA reference level and display to override the default setting and cycle through all
available selections: 5, 10, 15, 20, or 25 relative units, starting at the default
setting—until the button is released. Once the button is released, the UA trace and
UA value take on this new value as a baseline for reference.
Briefly pressing the
setting configured via the service setup mode.
If pressure falls below 0 relative units (probably because the belt has loosened),
automatic UA referencing occurs and a new baseline reference is set at 0 relative
units.
170 Series Monitor
2003023-001
button, if there is insufficient range to provide at
UA Reference
button sets the baseline at the default setting—the
UA Reference
"170 Series Service Manual"
UA Reference
button for more than two seconds causes the
button
UA Reference
for more information.
reverts back to using the default
Revision D