Polymobil 10
Maintenance Instructions
Maintenance protocol
RXR8-120.105.01.05.02 is required for
these instructions
Print No.: RXR8-120.101.01.05.02
Replaces: RXR8-120.101.01.04.02
SP
© Siemens AG 2005
The reproduction, transmission or
use of this document is not permit-
ted without express written author-
ity. Offenders will be liable for
damages.
All
rights,
including
rights created by patent grant or
registration of a utility model
design,_are_
reserved.
English
Date of creation: 07.05
_or_

Advertisement

Table of Contents
loading

Summary of Contents for Siemens Polymobil 10

  • Page 1 Polymobil 10 Maintenance Instructions © Siemens AG 2005 The reproduction, transmission or use of this document is not permit- ted without express written author- Maintenance protocol ity. Offenders will be liable for damages. rights, including RXR8-120.105.01.05.02 is required for rights created by patent grant or...
  • Page 2 Assemblers and other persons who are not employed by or otherwise directly affiliated with or autho- rized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or one of its affiliates before attempting installation or service procedures.
  • Page 3: Table Of Contents

    Leakage current/report ....... . . 3 - 3 4 _______Changes to Previous Version_____________________________________ 4 - 1 Siemens AG RXR8-120.101.01 Page 3 of 4 Polymobil 10 Medical Solutions Rev. 05 07.05 CS PS 24...
  • Page 4 0 - 4 Table of Contents Page Polymobil 10 RXR8-120.101.01 Page 4 of 4 Siemens AG Rev. 05 07.05 CS PS 24 Medical Solutions...
  • Page 5: Required Documents

    28 70 061 • Optimol Longtime PD2 grease 0.71 oz (20 g) 34 91 271 or 2.20 lb (1 kg) 73 95 445 Siemens AG RXR8-120.101.01 Page 1 of 14 Polymobil 10 Medical Solutions Rev. 05 07.05 CS PS 24...
  • Page 6: Emphasized Texts

    NOTE NOTE is used to indicate information which explains the proper way to use devices or to carry out a process, i.e. which provides hints and tips. Polymobil 10 RXR8-120.101.01 Page 2 of 14 Siemens AG Rev. 05 07.05...
  • Page 7: Safety Information And Protective Measures

    - Release radiation for as short a time as possible. Checks in which radiation must be released are identified by the radiation warning symbol. Siemens AG RXR8-120.101.01 Page 3 of 14 Polymobil 10 Medical Solutions Rev. 05 07.05 CS PS 24...
  • Page 8 1 - 4 General information Fig. 1 Polymobil 10 RXR8-120.101.01 Page 4 of 14 Siemens AG Rev. 05 07.05 CS PS 24 Medical Solutions...
  • Page 9 Only connect the POLYMOBIL 10 to a power supply (wall outlet) whose installation corresponds to the requirements according to VDE 0107 or the national regulations. • Prior to any intervention, switch the POLYMOBIL 10 off using the power OFF switch on the control console, and disconnect the power plug. •...
  • Page 10: Description Of Abbreviations

    During the measurement, move the power cable and additional connection cables with an integrated protective conductor section by section to detect cable breaks. The protective conductor resistance must not exceed 0.2 Ohms. Polymobil 10 RXR8-120.101.01 Page 6 of 14 Siemens AG Rev.
  • Page 11 "Certificates". If values are newly determined, they are to be recorded as first measured values. The evaluation is omitted in this case. Siemens AG RXR8-120.101.01 Page 7 of 14 Polymobil 10 Medical Solutions Rev. 05 07.05 CS PS 24...
  • Page 12: Information On Measuring The System Leakage Current

    However, please note the minimum resolution of the leakage current measuring instru- ment and any additional manufacturer's data restricting the use of the measuring device. Polymobil 10 RXR8-120.101.01 Page 8 of 14 Siemens AG Rev.
  • Page 13 DIN VDE 0751-1/2001-10, Fig. C5 for protection class 1 = System 2 = Applied part, type B (if present) 3 = Measuring arrangement (integrated in measuring instrument) Siemens AG RXR8-120.101.01 Page 9 of 14 Polymobil 10 Medical Solutions Rev. 05 07.05 CS PS 24...
  • Page 14 Separate this report, fill it out, and file it in the "Certificates" register in the system manual. If values are newly determined, they are to be recorded as first measured values. The evaluation is omitted in this case. Polymobil 10 RXR8-120.101.01 Page 10 of 14 Siemens AG Rev.
  • Page 15: Technical Safety Checks (Tsc)

    2.9 Radiation; tests: the devices present and not damaged? SIE Radiation indicator (no radiation protection of the building) SIE Acoustic signal Tab. 1 Siemens AG RXR8-120.101.01 Page 11 of 14 Polymobil 10 Medical Solutions Rev. 05 07.05 CS PS 24...
  • Page 16 SIM Support arm locking mechanism (lubricated, if required)? 2.6 Single tank; check: SIM Locking mechanism 2.7 Double-slot diaphragm, check: SIM Rotatability Tab. 1 Polymobil 10 RXR8-120.101.01 Page 12 of 14 Siemens AG Rev. 05 07.05 CS PS 24 Medical Solutions...
  • Page 17 (e.g., safety floor plate)? 5.7 Are the positions of the table and n.a. system reproducible (e.g., zero positions, layer height, etc.)? Tab. 1 Siemens AG RXR8-120.101.01 Page 13 of 14 Polymobil 10 Medical Solutions Rev. 05 07.05 CS PS 24...
  • Page 18 Have the operator, or a person authorized by the operator, sign the TSC report. Hand the TSC report over to the operator, or a person authorized by the operator. Tab. 1 Polymobil 10 RXR8-120.101.01 Page 14 of 14 Siemens AG Rev. 05 07.05...
  • Page 19: Visual Inspection

    Make sure that the handles are secure and tighten them if necessary: Single tank holder Single tank Control handle over the control console Siemens AG RXR8-120.101.01 Page 1 of 14 Polymobil 10 Medical Solutions Rev. 05 07.05 CS PS 24...
  • Page 20 2 - 2 Inspection and maintenance NOTE If screws or securing elements are replaced, the screw locks must also be replaced. Polymobil 10 RXR8-120.101.01 Page 2 of 14 Siemens AG Rev. 05 07.05 CS PS 24 Medical Solutions...
  • Page 21: Checking The Wheels

    • Pedal positions as shown in 3/Fig. 1 The unit can be moved forward or backward. It is steered through the front wheels. Siemens AG RXR8-120.101.01 Page 3 of 14 Polymobil 10 Medical Solutions Rev. 05 07.05 CS PS 24...
  • Page 22: Single Tank

    - Remove the clamp handle screw (4/Fig.3), and use the screwdriver to press the clamp handle away from the axle. - Attach the clamp handle in the vertical position and tighten the screw (4/Fig.3). - Recheck the locking mechanism of the single tank. Polymobil 10 RXR8-120.101.01 Page 4 of 14 Siemens AG Rev.
  • Page 23: Double-Slot Diaphragm

    Move the support arm to the top position. • Use the spirit level on the double-slot diaphragm to check the 0° position. If necessary, readjust the scale on both sides. Siemens AG RXR8-120.101.01 Page 5 of 14 Polymobil 10 Medical Solutions Rev. 05 07.05 CS PS 24...
  • Page 24: Stand

    Check the edges of the locking disks (6/Fig.6). They must not be worn. • Ensure that there is play (without any friction) on all sides of the locking bar in the stand head (7/Fig.6). Polymobil 10 RXR8-120.101.01 Page 6 of 14 Siemens AG Rev.
  • Page 25 Remove cover (3/Fig.7) from the stand. • Carefully check the condition of the spring counterweight chains (5/Fig.8). Replace the stand column if the chains are defective. Siemens AG RXR8-120.101.01 Page 7 of 14 Polymobil 10 Medical Solutions Rev. 05 07.05 CS PS 24...
  • Page 26: Radiation

    Trigger exposure. The radiation indicator (1/Fig.9) must be lit during exposure, and an acoustic signal will sound at the same time. • POLYMOBIL OFF and disconnect the power plug. Polymobil 10 RXR8-120.101.01 Page 8 of 14 Siemens AG Rev. 05 07.05...
  • Page 27 If the oscillogram ist not produced, perform an adjustment according to the service instructions. • POLYMOBIL OFF and disconnect the power plug. • Close the control console. Siemens AG RXR8-120.101.01 Page 9 of 14 Polymobil 10 Medical Solutions Rev. 05 07.05 CS PS 24...
  • Page 28 • Make a note on the processed film of the following data using a waterproof felt tip pen: - Adjusted SID - Film size - Radiation field size Polymobil 10 RXR8-120.101.01 Page 10 of 14 Siemens AG Rev. 05 07.05...
  • Page 29: Control Console

    Check the operating elements and indicators on the control console. - See "Overview of the Operating Elements" in the operator manual. • POLYMOBIL OFF and disconnect the power plug. Siemens AG RXR8-120.101.01 Page 11 of 14 Polymobil 10 Medical Solutions Rev. 05 07.05 CS PS 24...
  • Page 30: Protective Conductor Test

    Separate this report, fill it out, and file it in the system manual under "Certificates". If values are newly determined, they are to be recorded as first measured values. The evaluation is omitted in this case. Polymobil 10 RXR8-120.101.01 Page 12 of 14 Siemens AG Rev.
  • Page 31: Leakage Current Measurement

    "Certificates" register in the system manual. If values are newly determined, they are to be recorded as first measured values. The evaluation is omitted in this case. Siemens AG RXR8-120.101.01 Page 13 of 14 Polymobil 10 Medical Solutions Rev. 05 07.05 CS PS 24...
  • Page 32: Concluding Work

    Do not use scouring cleaning agents or organic solvents or solvent-based cleaning agents (for example, benzine, alcohol, spot remover) because of potential incompatibility with the material! Do not spray the POLYMOBIL! Never let cleaning agents get into the POLYMOBIL! Polymobil 10 RXR8-120.101.01 Page 14 of 14 Siemens AG Rev.
  • Page 33: Protective Conductor Resistance/Report

    Meas. inst. Ser. No.: Meas. inst. calibrated up to: n.a. Assessment: Date: Name: Signature: (*1) Meas. circuit: see Fig. 1, next page Siemens AG RXR8-120.101.01 Page 1 of 6 Polymobil 10 Medical Solutions Rev. 05 07.05 CS PS 24...
  • Page 34 DIN VDE 0751-1:2001-10, Fig. C2. 1 = Measuring arrangement (measuring instrument) 2 = System 3 = Application part (if present) Comments: Date Remarks Name Signature Polymobil 10 RXR8-120.101.01 Page 2 of 6 Siemens AG Rev. 05 07.05 CS PS 24 Medical Solutions...
  • Page 35: Leakage Current/Report

    Meas. inst. Ser. No.: Meas. inst. calibrated up to: Assessment: Date: Name: Signature: (*1) Meas. circuit: See Fig. 2 to Fig. 3, next page Siemens AG RXR8-120.101.01 Page 3 of 6 Polymobil 10 Medical Solutions Rev. 05 07.05 CS PS 24...
  • Page 36 DIN VDE 0751-1:2001-10, Fig. C6 for protective class 1 = System 2 = Application part (if present) 3 = Measuring arrangement (integrated in the measuring instrument) Polymobil 10 RXR8-120.101.01 Page 4 of 6 Siemens AG Rev. 05 07.05...
  • Page 37 Electrical safety/reports 3 - 5 Comments: Date Remarks Name Signature Siemens AG RXR8-120.101.01 Page 5 of 6 Polymobil 10 Medical Solutions Rev. 05 07.05 CS PS 24...
  • Page 38 3 - 6 Electrical safety/reports Date Remarks Name Signature Polymobil 10 RXR8-120.101.01 Page 6 of 6 Siemens AG Rev. 05 07.05 CS PS 24 Medical Solutions...
  • Page 39 Protective conductor test Completely rewritten maintenance Inspection and Leakage current Completely rewritten maintenance measurement Inspection and Concluding work "Disinfection" checkpoint deleted maintenance Electrical n.a. Chapter added safety/reports Siemens AG RXR8-120.101.01 Page 1 of 2 Polymobil 10 Medical Solutions Rev. 05 07.05 CS PS 24...
  • Page 40 4 - 2 Changes to Previous Version Polymobil 10 RXR8-120.101.01 Page 2 of 2 Siemens AG Rev. 05 07.05 CS PS 24 Medical Solutions...
  • Page 41 Address: Department: Room: Contact person: Telephone: Customer number: CS No.: Date: © Siemens AG 2005 The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders will be liable for damages. All rights, including rights...
  • Page 42 The Maintenance Protocol is a document and must therefore be archived. After comple- ting maintenance, it must be filed in the appropriate register of the Maintenance binder. If needed, give the customer a copy of it. Polymobil 10 RXR8-120.105.01 Page 2 of 8 Siemens AG Rev.
  • Page 43 For safety reasons, continued operation of the system is possible only after the deficiencies have been successfully corrected. Town: Date: Technician's name: Signature: Siemens AG RXR8-120.105.01 Page 3 of 8 Polymobil 10 Medical Solutions Rev. 05 07.05 CS PS 24...
  • Page 44 CS No.: Date: Activities performed If there is no entry in the table, see the entries beginning on Page 7. Date: Activities performed: not OK n.a. Polymobil 10 RXR8-120.105.01 Page 4 of 8 Siemens AG Rev. 05 07.05 CS PS 24...
  • Page 45 Preventive Maintenance, Preventive Parts Replacement, Visual Inspection Preventive maintenance adjustments Preventive Check of Operating Values/Functions Quality Check Image Quality Check System Quality Software Maintenance Customer Service Engineer Siemens AG RXR8-120.105.01 Page 5 of 8 Polymobil 10 Medical Solutions Rev. 05 07.05 CS PS 24...
  • Page 46 If 'Yes', enter the component (including Serial No.) into the table that was used for mainte- nance. Movable components (including test phantoms that are part of the system) are parts that can be used on different systems. Movable Component Serial No. Polymobil 10 RXR8-120.105.01 Page 6 of 8 Siemens AG Rev. 05 07.05...
  • Page 47 2 - 5 Light localizer lamp 2 - 5 Rotatability 2 - 5 Format setting 2 - 5 0° position 2 - 5 Siemens AG RXR8-120.105.01 Page 7 of 8 Polymobil 10 Medical Solutions Rev. 05 07.05 CS PS 24...
  • Page 48 3 - 1 3.1. Protective conductor resistance/report 3 - 1 3.2. System leakage current/report 3 - 3 Changes to previous version 4 - 1 Polymobil 10 RXR8-120.105.01 Page 8 of 8 Siemens AG Rev. 05 07.05 CS PS 24 Medical Solutions...

Table of Contents