LRF 6019 Integrator Manual
6.5.8
Continuous Range Request >LM,Mt,x*<CS><CR>
This command initiates a range request with predefined repetition rate.
The parameter x can take the following values:
x=0 Continuous range STOP.
x=1 Repetition rate of 0.5 Hz.
x=2 Repetition rate of 1 Hz.
x=3 Repetition rate of 3 Hz.
x=4 Repetition rate of 5 Hz.
x=5 Repetition rate of 10 Hz.
Continuous range command and response
Command
Inputs
>LM,Mt,x*<CS><CR>
>LM,Mt,5*E7<0D>
Response
Outputs
>LM,Mt,range1,range2,range3,timestamp*<CS><CR>
Every 0.1 seconds...
>LM,Mt,v0004252,R000E301,R000E301*53<0D>
Example of a successful 10 Hz continuous range request command transaction:
Host LRF Module
Command frame:
LRF Module Host
Acknowledge frame:
Response frame:
Host LRF Module
Command frame:
Acknowledge frame:
Prompt:
914928_TML_LRF6019_en_Version C
Confidential & Proprietary Safran Vectronix AG – All rights reserved
Table 10: 10 Hz continuous range request command
>LM,Mt,5*E7<0D>
>AC*84<0D>
>LM,Mt,v0006404,v0090656,v0077726,t018692980*1C<0D>
>LM,Mt,v0006406,v0090981,v0087450,t018758928*1F<0D>
>LM,Mt,v0072447,v0085329,v0074191,t018862206*16<0D>
>...
>LM,Mt,0*E2<0D>
>AC*84<0D>
<
Description
Continuous range request with parameter x repetition rate.
Example with x=5, repetition rate of 10Hz.
Description
Response of LRF module to the host with parameter x.
Example with x=5, repetition rate of 10 Hz.
Public: 04.2019
Page: 48/70