Recognizing module: 9 = Erbe Irrigation Pump EIP 2
Recog-
Error
nizing
code
module
9
80
9
82
9
84
9
88
9
90
9
A0
9
C0

9-Errors

Additional infor-
Description
mation
CRC Error.
Pump lid open at
User error:
ON.
The pump lid of the EIP 2 is
not closed at the start of acti-
vation or is opened during
activation.
Unit error.
Time Out Error.
User error:
Activation of the EIP 2 was
maintained for more than
20 seconds. Maximum activa-
tion time for the EIP 2 is
20 seconds.
UeWS Error
No current or no
load.
Current too high at
ON.
Motor control faulty
(voltage too high at
ON).
5 • Remedying malfunctions
Action
Restart the unit. If the error
occurs again, notify Erbe Ser-
vice.
Close the pump lid before
activation or do not open it
during activation.
Restart the unit. If the error
occurs again, notify Erbe Ser-
vice.
End the activation. It is then
possible to active the EIP 2
new. If the error occurs again,
check the accessories.
Restart the unit. If the error
occurs again, notify Erbe Ser-
vice.
Restart the unit. If the error
occurs again, notify Erbe Ser-
vice.
Restart the unit. If the error
occurs again, notify Erbe Ser-
vice.
Restart the unit. If the error
occurs again, notify Erbe Ser-
vice.
37