Test procedure VIO 200 S
1. Set test specimen to:
2. Set HF power meter to:
3. Activate test specimen via CUT button on the electrode handle.
4. Determine and document measured value. The tolerance range is 160
Test procedure VIO 300 S
1. Set test specimen to:
2. Set HF power meter to:
3. Activate test specimen via CUT button on the electrode handle.
4. Determine and document measured value. The tolerance range is 240
1. Set test specimen to:
2. Set HF power meter to:
3. Activate test specimen via CUT button on the electrode handle.
4. Determine and document measured value. The tolerance range is 160
BIPOLAR CUT
Test setup
(depending on the
unit variant)
Fig. 6-6
• The test specimen is connected to the power supply via the power
• The levels of power are determined with the HF power meter. The
• The dual-pedal footswitch is connected.
AUTO CUT, Effect 8, 200 watts
RL = 500 ohms
to 240 watts.
AUTO CUT, Effect 8, 300 watts
RL = 500 ohms
to 360 watts.
DRY CUT, Effect 8, 200 watts
RL = 500 ohms
to 240 watts.
+
+
cord.
measuring cables are plugged into the HF power meter direct.
6 • Maintenance and servicing
+
+
Bipolar
HF power meter
53
/ 64