QUICKSTART GUIDE for VIO
Set up for Snare/Hot Biopsy Use
1
Turn ON the Unit.
2
The message box will say, "Please confirm the
settings by pressing any key!" (Note: the last
program used will be the one displayed). To
choose a different program: Prog. Setup → Select
program → scroll to Snare/Hot Biopsy.
3
Press the FOCUS Active Display button beside the
Monopolar receptacle.
VIO
200 S
®
200 S
®
4
Check the Program Screen to confirm the settings
(Default program shown).
5
Plug in the Active Cord and attach the Snare or
Hot Biopsy Forceps.
6
Plug the Grounding Pad into the NEUTRAL
receptacle and place the Pad on the patient.
7
Place Footswitch at the physician's feet. Verbally
confirm settings with physician.
8
READY to begin procedure.
Adjust settings per physician direction, increasing the
Effect then max. Watts for more power.
LIT/5213/02 02/17

Advertisement

loading
Need help?

Need help?

Do you have a question about the VIO 200 S and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Erbe VIO 200 S

  • Page 1 QUICKSTART GUIDE for VIO 200 S ® Set up for Snare/Hot Biopsy Use Check the Program Screen to confirm the settings Turn ON the Unit. (Default program shown). The message box will say, “Please confirm the settings by pressing any key!” (Note: the last program used will be the one displayed).
  • Page 2 QUICKSTART GUIDE for VIO 200 S ® Set up for Bipolar Use Check the Program Screen to confirm the settings Turn ON the Unit. (Default program shown). The message box will say, “Please confirm the settings by pressing any key!” (Note: the last program used will be the one displayed).