5 • Remedying malfunctions
Recognizing module: A = APC 2, E = Extension module VEM 2
Recog-
Error
nizing
code
module
A/E
7D + 7E
A/E
7F
A/E
80
A/E
81 – 83
A/E
85
A
86
A
90
A
91
A
A0
24
/ 64
Additional infor-
mation
Description
Operating system error.
Internal state incorrect.
Protocol violation CAN.
Invalid resistance instrument
number.
Gas underdose, e.g. hose
blocked.
Low pressure at cylinder 1.
Low pressure at cylinder 2.
Action
Restart the unit. If the error
occurs again, notify Erbe Ser-
vice.
Information, no fault condi-
tion.
Information, no fault condi-
tion.
Restart the unit. If the error
occurs again, notify Erbe Ser-
vice.
Check the accessories. If the
error persists, notify Erbe
Service.
Check the accessories. If the
error persists, notify Erbe
Service.
Check the gas supply (tank,
pressure regulator). If the
error persists, notify Erbe
Service.
Check the gas supply (tank,
pressure regulator). If the
error persists, notify Erbe
Service.
Restart the unit. If the error
occurs again, notify Erbe Ser-
vice.