Recognizing module: B = Control panel
Recog-
Error
nizing
code
module
B
84
B
85
B
87
B
8B
B
8D
B
8E + 8F
B
90
B
91
B
95
B
9F
Additional infor-
Description
mation
Footswitch was recognized
by the system:
Occurs if a footswitch is con-
nected.
Footswitch was discon-
nected from the system:
Occurs if a footswitch is dis-
connected from the system.
No status message from APC
module:
Occurs if an APC module is
disconnected from the sys-
tem.
No status message from
IES module:
Occurs if the IES module is
disconnected from the sys-
tem.
New instrument recognized
by system:
Occurs if an instrument with
instrument recognition is
connected.
Instrument has been discon-
nected from the system:
Occurs if an instrument with
instrument recognition is
disconnected from the sys-
tem.
5 • Remedying malfunctions
Action
Information, no fault condi-
tion.
Information, no fault condi-
tion.
Restart the unit. If the error
occurs again, notify Erbe Ser-
vice.
Restart the unit. If the error
occurs again, notify Erbe Ser-
vice.
Check connectors. Check
APC.
Restart the unit. If the error
occurs again, notify Erbe Ser-
vice.
Restart the unit. If the error
occurs again, notify Erbe Ser-
vice.
Check connectors. Check
IES 2.
Information, no fault condi-
tion.
Information, no fault condi-
tion.