5 • Remedying malfunctions
Recognizing module: D = Smoke evacuation system IES 2
Recog-
Error
nizing
code
module
D
80
D
81
D
82
D
83
D
86
D
87
D
88
32
/ 64
D-Errors
Additional infor-
mation
Description
ECB fault.
The VIO first receives a one-
second signal from the IES 2
and then from the VIO master
unit. The VIO interprets this
condition as the availability
of two master units. The con-
flict is resolved without inter-
vention once the IES 2 has
entered slave operation.
Excess temperature
(> 60 °C).
Insufficient temperature
(< 0 °C).
Obstruction.
The filter is missing or
obstructed.
Action
Restart the unit. If the error
occurs again, notify Erbe Ser-
vice.
Information, no fault condi-
tion.
Change the activation
sequence of the units. First
switch on the VIO followed by
the IES 2.
Overtemperature IES2: Let
the unit cool down.
Undertemperature IES2:
Leave unit on for several
minutes without operating.
Restart the unit. If the error
occurs again, notify Erbe Ser-
vice.
Restart the unit. If the error
occurs again, notify Erbe Ser-
vice.
Remove the suction tube
from the tissue.
Insert or replace the filter.