Recognizing module: B = Control panel
Recog-
Error
nizing
code
module
B
C6
B
D2
B
D3
B
D4 – D6
B
DA
B
DB
B
DC
B
EA
B
EB
Additional infor-
Description
mation
Temperature monitoring for
neutral electrodes: NESSY
current > 300 mA
No status message from VEM
module:
Occurs if the VEM module is
disconnected from the sys-
tem.
No status message from EIP
module:
Occurs if the EIP module is
disconnected from the sys-
tem.
Occurs if a module responds
that is not supported by the
system.
Occurs if a button that is not
selectable is pressed (after a
timeout).
Module was recognized:
Occurs if a module is recog-
nized.
Receptacle was recognized:
Occurs if a receptacle is rec-
ognized.
5 • Remedying malfunctions
Action
Information, no fault condi-
tion.
Information, no fault condi-
tion.
Information, no fault condi-
tion.
Restart the unit. If the error
occurs again, notify Erbe Ser-
vice.
Remove the module con-
cerned from the unit and
notify Erbe Service.
Remove the module con-
cerned from the unit and
notify Erbe Service.
Restart the unit. If the error
occurs again, notify Erbe Ser-
vice.
Information, no fault condi-
tion.
Check and save the recepta-
cle configuration in the "Ver-
sion List" test program. If the
error occurs again, notify
Erbe Service.