Err = Error display on the seven-segment displays
Display
Description
ErrdIS
ErrtAS
Recognizing module: F = Footswitch
Recog-
Error
nizing
code
module
F
01
Err-Errors
F-Errors
Additional infor-
Description
mation
CRC error:
Occurs if an error is detected
during CRC monitoring of the
footswitch at start-up (e.g.
bit failure in the Flash mem-
ory or error when download-
ing software).
5 • Remedying malfunctions
Action
Restart the unit. If the error occurs again,
notify Erbe Service.
Restart the unit. If the error occurs again,
notify Erbe Service.
Action
Replace the ECB footswitch.