Service And Error Messages In The Stand-By Menu - KaVo ESTETICA E30 Short Instructions For Use

Hide thumbs Also See for ESTETICA E30:
Table of Contents

Advertisement

Short instructions for use 9 Troubleshooting / warning signals
Malfunction
The suction hoses do
not have any suction.
Water in the return
air filter.
The operating light
cannot be switched
on.

9.3 Service and error messages in the stand-by menu

Service and error messages are displayed in the stand-by menu through the "!" symbol.
S1
S2
S3
▶ Press the "S2" button to display messages.
▶ Press the function keys to navigate between the messages.
Key
Settings
S2
Previous message
S3
Next message
S4
Switch to stand-by menu
Cause
Slides on the conical
sections are closed.
Sieves in suction
connector are
blocked.
Base plate for vacu-
stop has been activ-
ated.
Suction machine not
running.
O-rings of the MUL-
TIflex coupling are
damaged.
Switch on the lamp
head is switched off.
S4
S1
Remedy
▶ Open the slide valve.
▶ Replace sieves.
▶ Relieve base plate.
▶ Turn on the suction machine.
▶ Check the suction machine fuse.
▶ Replace all O-rings of the MULTIflex
coupling.
▶ Turn on the switch.
S2
S3
S4
30