3 Preparation
3.2.2 Removing the DUIS Focus Unit ASSY
1
Remove the ionization unit.
2
Remove the ground plate.
3
Remove the DUIS focus unit ASSY.
28
™
IonFocus
Unit
Check that the temperature of the DL, heat block, and interface heater has
fallen to 50 °C or lower before proceeding with the work.
The inside of the ionization unit and the area around the DL and heat block
reach high temperatures.
Turn the heater OFF from the LabSolutions before starting this work.
If you proceed with the work while the temperature is still high, you may
sustain burns.
"LCMS Instruction Manual", "Heater ON/OFF Status and Temperature
Monitor"
Take care because the tip of the needle unit is sharp.
"3.1.6 Removing the Ionization Unit" P.21