If it is not possible to use a conductive waste container, take the following precautions:
Ensure that the end of the inflow tubing is always submerged inside the container. Also,
place some type of grounded metal object, such as a ground wire connected to the
instrument, into the liquid.
Note, however, that the above precaution will be ineffective for low conductivity (less
-10
than 10
S/m) liquids.
Keep the room at a proper humidity.
Ambient humidity exceeding 65 % will prevent static electricity.
Use as small a container as possible to minimize damage in the event of fire.
This minimizes the damage in the event of fire.
<Reference>
Anti-static equipment (anti-static clothing, shoes and matting) and charge measurement
equipment (electrometer) are sold by specialty manufacturers.