Endress+Hauser analytikjena multi N/C UVHS Operating Manual page 95

Toc analyzer
Table of Contents

Advertisement

multi N/C UVHS, multi N/C pharma UV
Error
Cause
¡
The measuring gas flow is below the
range of 140 ml/min.
Error
Cause
The measuring gas flow drops below the
¡
minimum flow during measurement
pauses. The UV lamps switches off for
safety reasons.
¡
Lamp does not switch on despite correct
gas flow. UV lamp defective
Error
Cause
¡
Selected sample volume too large.
¡
Number of measurements too high.
Error
Cause
¡
Flushing reservoir insufficient.
Error
Cause
Dosing faulty.
¡
The dosing syringe is leaky.
¡
¡
Instable reagent addition.
Inhomogeneous samples.
¡
¡
Sensitive samples can be affected by
ambient air.
NDIR-based drift:
¡
Unsuitable integration criteria.
The software ends measurement too
early.
UV lamp does not switch on during initial-
ization
Remedy
¡
Check the gas flow, see Status errors.
UV lamp switches off during operation
Remedy
Check the gas flow, see Status errors.
¡
¡
Inform customer service and have the
UV module replaced.
Minimum sample volume > vessel volume
Remedy
¡
Check and adjust the sample volume,
flushing volume and number of determi-
nations settings in the method.
Flushing water insufficient (for sample feed
with autosampler)
Remedy
¡
Check and adjust the sample volume and
backwash settings in the method.
Scattering measurements
Remedy
Check the dosing.
¡
Ensure that samples are drawn in with-
¡
out bubbles.
¡
Check that enough sample is provided.
¡
Install a new dosing syringe.
¡
Ensure that reagents are drawn in with-
out bubbles.
¡
Take the reagent blank value into ac-
count.
¡
Adjust the purge flow for reagents.
Check that sufficient reagent is provided.
¡
Insert the hose further into the reservoir
¡
bottle.
Filter samples prior to analysis.
¡
Stir samples prior to injection. To do
¡
this, use an autosampler with stirring
function.
¡
Prevent the addition of CO
vapors from the ambient air.
¡
Check the ambient conditions and rem-
edy the source of the fault.
Cover the sample vessels on the au-
¡
tosampler with aluminum foil.
Treat the headspace of the sample with
¡
gas.
Check the method settings.
¡
¡
If necessary, increase the maximum in-
tegration time.
Troubleshooting
or organic
2
95