multi N/C UVHS, multi N/C pharma UV
Error code: Error message
Cause
¡
The analog values of the detector are
outside of the working range.
Error code: Error message
Cause
¡
No communication between the ana-
lyzer and the syringe pump.
Error code: Error message
Cause
No connection to the solids module.
¡
¡
The solids module is not switched on.
¡
Incorrect connection.
Error code: Error message
Cause
Faulty thermocouple.
¡
¡
Furnace not connected.
¡
The furnace temperature is too high
Error code: Error message
Cause
¡
The contact on the UV cover is not
closed, for example after the UV module
has been replaced.
Error code: Error message
Cause
¡
Communication error.
Error code: Error message
Cause
No connection to the solids module.
¡
Error code: Error message
Cause
The drive is incorrectly positioned, e.g.
¡
jammed.
The drive is faulty.
¡
Error code: Error message
Cause
¡
The drive is incorrectly positioned, e.g.
jammed.
The drive is faulty.
¡
27: Optics error, analog values out of range
Remedy
¡
Check the quality of the carrier gas.
¡
For solids methods and connection of
the HT 1300 module: Set the carrier gas
flow higher than the intake flow.
¡
Initialize the analyzer and check the
analog values via the component test.
40: No connection to the syringe pump
Remedy
¡
Initialize the analyzer.
¡
Switch the PC off and back on again and
initialize the analyzer.
80: No connection to temperature controller
Remedy
Check the connection cables.
¡
¡
Switch on the optional solids module.
¡
Check the connection.
81: Thermocouple HT furnace interruption
(HT) / UV cover open (UV)
Remedy
Inform the service.
¡
¡
Connect the furnace.
¡
Inform the service.
81: Thermocouple HT furnace interruption
(HT) / UV cover open (UV)
Remedy
¡
Close the cover.
84: Communication error HT furnace tem-
perature controller
Remedy
¡
Inform the service.
86: No external furnace found
Remedy
Check the connection cables.
¡
111: Rotator error
Remedy
Initialize the analyzer.
¡
If the error cannot be corrected, contact
¡
the service.
112: Swivel drive error
Remedy
¡
Initialize the analyzer.
¡
If the error cannot be corrected, contact
the service.
Troubleshooting
91