Troubleshooting
92
Error code: Error message
Cause
¡
The drive is incorrectly positioned, e.g.
jammed.
¡
The drive is faulty.
Error code: Error message
Cause
The sample tray is not positioned cor-
¡
rectly.
Error code: Error message
Cause
The wrong sample tray is set in the soft-
¡
ware.
Error code: Error message
Cause
Communication error.
¡
Error code: Error message
Cause
¡
Internal computer reset.
¡
Overvoltage.
¡
Short-term power failure.
Error code: Error message
Cause
¡
Internal program error.
Error code: Error message
Cause
Program error.
¡
Error code: Error message
Cause
¡
Communication error.
¡
Syringe pump faulty.
Error code: Error message
Cause
The analyzer is not yet initialized after
¡
switching on.
¡
Reset of the syringe pump.
multi N/C UVHS, multi N/C pharma UV
113: Lifting drive error / Sampler: z drive er-
ror (steps lost)
Remedy
¡
Initialize the analyzer.
¡
If the error cannot be corrected, contact
the service.
114: Rack detection error
Remedy
Position the sample tray again and make
¡
sure it clicks into place.
¡
Initialize the analyzer.
115: Wrong rack
Remedy
Check the software settings.
¡
If necessary, set a different sample tray.
¡
116: Unknown sampler command
Remedy
Inform the service.
¡
200: Restart computer in the analyzer
Remedy
¡
If the LED status display is lit, initialize
the analyzer.
¡
For repeated occurrences, monitor pre-
cisely at which time the error occurs. Ob-
serve the status bar.
201: Restart the internal program
Remedy
¡
Initialize the analyzer.
¡
For repeated occurrences, monitor pre-
cisely at which time the error occurs. Ob-
serve the status bar!
202: File method.txt not found
203: File init.cnf not found
Remedy
Inform the service.
¡
401: Syringe pump: Initialization
402: Syringe pump: invalid command
403: Syringe pump: invalid operand
404: Syringe pump: faulty command se-
quence
Remedy
¡
Initialize the analyzer.
¡
Inform the service.
407: Syringe pump: syringe pump not initial-
ized
Remedy
Initialize the analyzer.
¡

Advertisement