Installation and commissioning
4.3.1.1 Installing and commissioning the sampler
42
CAUTION
Risk of injury from moving parts
There is a risk of injury in the movement range of the sampler arm. For example, hands
or fingers might be crushed.
¡ Maintain a safety distance from the sampler during operation.
NOTICE
Risk of device damage
If the sampler arm is obstructed during operation, the drives can be destroyed.
¡ Do not touch the sampler arm during operation.
¡ Only carry out manual adjustment when the device is switched off.
} Switch off the analyzer before installing the autosampler.
} Plug the grounding conductor into the connection on the rear of the analyzer. Con-
nect the grounding conductor to the connection on the bottom of the sampler.
} Plug the cable on the low voltage side of the external power supply unit into the con-
nection on the bottom of the sampler. Do not connect the power supply unit to the
mains power supply yet.
} Connect the autosampler to the analyzer with the interface cable (interface on the
bottom of the sampler and "sampler" interface on the rear of the analyzer).
Figure 21Connections on the bottom of the autosampler
1 Connection for equipotential bonding
cable (grounding cable)
3 Analyzer interface
multi N/C UVHS, multi N/C pharma UV
1
2
3
2 Power cable connection