Replacing The Loader Stepper Assembly; Tools Required; Removing The Loader Stepper Assembly; Installing The Loader Stepper Assembly - Kodak CR 360 Service Manual

Table of Contents

Advertisement

Replacing the Loader Stepper Assembly

Tools Required

• Cassette Loader Kit
• Long-shaft 2.5 mm Allen wrench
• 2.5 mm x 100 mm cable ties
• Wire cutters
• Tape

Removing the Loader Stepper Assembly

[1] Disconnect the scanner from the main power.
[2] Remove the scanner cover. (See
[3] Remove the roof assembly and the tray assembly. (See
Removing the Tray Assembly on Page
[4] Disconnect the stepper motor cable CB057022.
[5] Disconnect sensor CB057040 from J508 on the connector panel.
[6] Remove the 4 screws that attach the loader stepper assembly to the loader frame using the 2.5 mm Allen
wrench.
Caution
Be sure to remove the cables carefully.
[7] Pull the loader stepper assembly towards the front of the unit and carefully remove it from the unit.

Installing the Loader Stepper Assembly

[1] Pass the stepper motor cables under the drum.
[2] Install the two Allen screws in the back.
[3] Install the loader stepper assembly into the motor frame.
[4] Insert but do not tighten the front two 2.5 mm screws.
[5] Connect the sensor cable to the sensor board and the motion cable to the motion board and fasten the cable ties.
[6] Adjust the loader pusher position. (See
[7] Before replacing the scanner cover:
(a) Go to Setup>Diagnostic>Motion.
(b) Operate the rollers forward and backward (Load/unload).
(c) Load a cassette into the scanner and check that the scanner loads and unloads the cassette correctly.
6H4869
Removing the Scanner Cover on Page
Removing the Roof Assembly on Page 83
85.)
Adjustment Procedure on Page
Service Procedures
25.)
129.)
and
67