Installing the Right Limit Sensor

[1] Install the right limit sensor into the sensor board.
[2] Position the sensor on the sensor bracket according to the marking made during the removal procedure.
[3] Install the two 2.0 mm retaining screws.
Note
The sensor bracket head must be adjusted so the sensor screen is in the middle of the sensor bracket, with equal
spacing between the two sides of the sensor screen and the bracket.
Laser Warning
When a service operation is taking place with the cover removed, disconnect the laser according to the procedure in
Activating the Scanner Without the Cover (Laser Deactivated) on Page
service procedure, wear protective safety glasses at all times.
[4] Before replacing the scanner cover:
(a) Go to Settings>Diagnostics>Motion.
(b) In the Linear Motor section, click << or >> to activate the linear motor.
(c) The carrier begins to move, and should enter the sensor area causing the indicator light to illuminate.
[5] Install the scanner cover.

Removing the Left Limit Sensor

[1] Disconnect the scanner from the main power.
[2] Remove the scanner cover. (See
[3] Mark the position of the sensor on the bracket.
[4] Disconnect the left limit sensor cable from the sensor board.
[5] Remove the two 2.0 mm retaining screws from the sensor.
[6] Remove the sensor from the sensor bracket.
6H4869
Removing the Scanner Cover on Page
26. If the laser must be activated during the
25.)
Service Procedures