Replacing the Loader Stepper Motor

Introduction
Laser Warning
When a service operation is taking place with the cover removed, disconnect the Laser according to the procedure
in
"Disconnecting the Laser" on Page
safety glasses at all times.
The Loader Stepper Motor Assembly unloads the plate from the Cassette for insertion into the Drum.

Stepper Motor

Stepper Motor
Carriage
Stepper Motor
Unit
Tools Required
• Replacement Stepper Motor
• 3.0 mm Allen wrench
• 2.5 mm x 100 mm cable ties
• Cutter
• Tray Gauge (T00900251)
• Feeler Gauge 0.5 mm
• Feeler Gauge 0.8 mm
Removing the Stepper Motor
[1] Disconnect the Scanner from the main power.
[2] Remove the Service Panel and Scanner Cover. (See
Catalog No. 6H1344 – 10APR2006
35. If the Laser must be activated during the service procedure, wear protective
"Removing the Main Cover" on Page 33
Replacing Components
3 mm Stepper Motor
retaining screws
.)
43