Cleaning The Module; Remove And Install Doors - Agilent Technologies InfinityLab LC Series User Manual

Diode array detectors
Hide thumbs Also See for InfinityLab LC Series:
Table of Contents

Advertisement

9
Maintenance

Cleaning the Module

Cleaning the Module

To keep the module case clean, use a soft cloth slightly dampened with water, or
a solution of water and mild detergent. Avoid using organic solvents for cleaning
purposes. They can cause damage to plastic parts.
Liquid dripping into the electronic compartment of your module can cause
WARN IN G
shock hazard and damage the module
Do not use an excessively damp cloth during cleaning.
Drain all solvent lines before opening any connections in the flow path.
A solution of 70 % isopropanol and 30 % water might be used if the surface of the
NOT E
module needs to be disinfected.

Remove and Install Doors

Parts required
p/n
5067-5737
5067-5736
The figures shown in this procedure exemplarily show the Infinity II Multisampler
NOT E
module.
The principle of how to remove and/or install doors works in the same way for all
Infinity II modules.
Agilent InfinityLab LC Series Diode Array Detectors User Manual
Description
Door left
Door right
157