XK10 Hardware
XK10 Hardware
XK10 Software
Flatness
Level
Performance specifications continued
Parallelism
Range
Accuracy (i)
Accuracy (ii)
Resolution
* laser to pentaprism distance > 0.3 m
A = (largest) straightness reading (µm)
M = length of the axis (m)
i. To be used when the quantity of interest is the angle between rails.
ii. To be used when parallelism between rails is:
specified as a tolerance zone defined by two parallel lines parallel to
a datum axis (for example, reference rail) within where the axis of the
feature (for example, measurement rail) must lie.
intended as a point by point variation in the separation between the rails,
with respect to the separation between the first two points
Spindle direction
Range
Accuracy (vertical)
Accuracy (horizontal)
Resolution
XK10 alignment laser system
XK10 Applications
Straightness
Parallelism
Coaxiality
±5 mm
±0.01A/M ±2/M ±4 µm/m*
±0.01A ±2 ±4M µm*
0.1 µm
±5 mm
±3 µm / 300 mm
±1.5 µm / 300 mm
0.1 µm
Squareness
Spindle direction
Coaxiality (shaft alignment)
Range
Accuracy (angle)
Accuracy (offset)
Resolution
NOTE: To achieve the specified performance, the launch unit should
only be used with the S unit and M unit that it was originally paired with.
This information can be found on the calibration certificate which is supplied
with the XK10 system.
±5 mm
±1 µm / 100 mm
±1 µm
0.1 µm
www.renishaw.com/xk10