XK10 Hardware
XK10 Software
Flatness
Level
Visual alignment of the launch unit to the reference rail
18. Move the M unit to the furthest position on the structure.
19. Adjust
the yaw of
the launch unit
so that the beam is
in line with the centre
of the target.
XK10 alignment laser system
XK10 Applications
XK10 Applications
Straightness
Parallelism
Parallelism
Coaxiality
Squareness
Spindle direction
NOTE: It is likely that the yaw of the launch unit
will need to be adjusted so much that the beam no
longer enters the input aperture. Slide the mirror
back over the input aperture and translate the
pentaprism until the beam is back on the centre,
and then continue with the launch unit alignment.
www.renishaw.com/xk10