Other Display Messages - Abbott GEMSTAR Operating Manual

Table of Contents

Advertisement

Other Display Messages

Other Display Message
KEYPAD LOCKED
NOT ALLOWED
DURING INFUSION
PRESS STOP TO
HALT DELIVERY
ROUNDING
The pump sounds four quick
beeps and displays the rounded
number.
DOSE IN PROGRESS
NEW CONTAINER
NOT ALLOWED
CANNOT CHANGE
CLOCK WHILE TIME
BASED DELIVERY
IS IN PROGRESS
PROGRAM EXCEEDS
MAX CONTAINER
SIZE. CHANGE
PROGRAM VALUES.
THIS OPTION NOT
AVAILABLE WHILE
PROGRAMMING THE
PUMP
Abbott GemStar Manual 430-600095-002 (Rev. 8/02)
Cause/Condition
User has attempted to perform a
function that has been locked out.
User has attempted to perform a
function that is not available in
RUN mode.
The pump has rounded a number
that was entered during
programming.
User has attempted to access NEW
CONTAINER function when it is
not allowed.
User has attempted to change the
clock during a time-based delivery.
The VTBI + KVO volume resulted in
a container size greater than the
allowed value.
User has attempted to access a
function that is not available from
the first programming display.
Troubleshooting
Remedy
Unlock the keypad to
access the desired
function.
• Press [STOP] to
access the desired
function.
• Press [YES/ENTER]
to return to the RUN
mode display.
Press [YES/ENTER] to
accept the rounded
value and continue to
the next step.
Press [STOP] to stop
delivery, then access
NEW CONTAINER
through the Change
Menu.
1. Press [STOP].
2. Clear the program
through the Change
Menu.
3. Change the clock
through the Options
Menu.
4. Reprogram the
pump.
1. Press [HELP] to
display the allowed
values.
2. Change the program
values.
1. Press [YES/ENTER]
to return to the first
programming
display.
2. Complete
programming.
3. Access desired
function.
97