Abbott GEMSTAR Operating Manual page 32

Table of Contents

Advertisement

8
Push down on the flow stop
is in the closed position.
9
Insert the cassette into the cassette pocket. Refer to "Loading the Cassette" on page
26.
d
Note: Extension set with anti-siphon valve must be attached before patient-use.
10
Connect the extension set to the pump set (refer to the instructions that come with
the pump set).
11
Press [
PURGE
d
12
Press and hold [
Note: Purge rate is 250 mL/hr. The pump purges for up to 2 minutes each time you press
and hold [
PURGE
Continue using the PURGE key until fluid fills the tubing. Ensure all air is removed
from the container, cassette, and tubing.
CAUTION:
When priming is complete, ensure no fluid flows at the distal end of
the pump set. If flow is observed, DO NOT use the pump set.
13
If extension set has a Y-extension, prime it if needed.
14
Connect the pump set to the patient-access device.
Abbott GemStar Manual 430-600095-002 (Rev. 8/02)