Abbott GEMSTAR Operating Manual page 59

Table of Contents

Advertisement

Intermittent
For example, in the following diagram, an intermittent dose was started at 2:00. The
dose was interrupted during the current dose cycle and was started again before the
end of the current dose cycle.
Dose Delivered -
If an Intermittent dose is interrupted and the pump is placed in the
6
the off cycle of the current dose, the next dose begins delivery and all subsequent
doses are shifted if necessary.
For example, in the following diagram, an intermittent dose was stared at 2:00. It
was interrupted during the current delivery cycle at 2:30, but was not started again
until after the current delivery cycle at 6:45:
Dose Delivered -
When the dose is complete, the start time for the Next Dose displays. To change the
6
start time refer to the "Options Menu" section.
Any missed Intermittent doses, except the first dose, are added to the end of the
6
delivery schedule.
52
12
1
2
3
4
5
6
7
Dose Interrupted -
12
1
2
3
4
5
6
7
Dose Interrupted -
Abbott GemStar Manual 430-600095-002 (Rev. 8/02)
8
9
10
11
12
RUN
8
9
10
11
12
mode after