Abbott GEMSTAR Operating Manual page 102

Table of Contents

Advertisement

Alert Display Message
Y = Yes
N = No
CHECK PRINTER
EMPTY CONTAINER
The pump does not
infuse unless Auto
KVO is delivering in a
mL/hr Only or
Weight Dosed
programs.
Abbott GemStar Manual 430-600095-002 (Rev. 8/02)
Is the pump
Alarm
infusing?
Condition
A = Audible alarm
V = Visual alarm
Y
A/V
• The printer has not
responded to
commands from the
pump.
• The printer is not
properly connected
to the pump.
N
A/V
The pump has
delivered the
programmed container
size or the pump has
delivered the VTBI
programmed AND Auto
KVO is delivering.
Troubleshooting
Cause/
Remedy
1. Press [SILENCE] to
quiet the alarm for 2
minutes.
2. Ensure the printer is
properly connected to
the pump by checking
the following:
Þ Printer is powered
on and in ONLINE
mode.
Þ Cable is plugged in
to both the pump
and the printer.
Þ Cable part number
and printer settings
are correct.
3. If the problem
continues, select STOP
PRINT from the Print
Menu to cancel all print
jobs.
1. Press [SILENCE] to
quiet the alarm for 2
minutes.
2. Press [STOP] to place
the pump in the STOP
mode and clear the
alarm.
3. Use the NEW
CONTAINER function
to repeat the current
program. Refer to
Starting a New Container
in the "Basic
Operation" section.
95