Abbott GEMSTAR Operating Manual page 99

Table of Contents

Advertisement

Troubleshooting
Alarm Display Message
Y = Yes
N = No
LOW BATTERIES
POWER LOSS
92
Is the
pump
Alarm
Condition
infusing?
A = Audible alarm
V = Visual alarm
Y
A/V
The pump has detected
battery voltage below
acceptable levels and the
batteries are the primary
power source.
Note: The pump allows the
user to clear the LOW
BATTERIES audible and
visual alarm. If the condition
has not been corrected after
15 minutes, the audible and
visual alarms resume.
If the pump is infusing and a
LOW BATTERIES alarm
occurs, the pump continues
for approximately 30
minutes* before it can no
longer deliver due to
insufficient power.
*If an external power source
is used, the rate is changed,
or the pump is powered-on
after a LOW BATTERIES
alarm and those same
batteries are still in the
pump, this interval may be
different.
N
A/V
The pump has detected a
power failure due to one
of the following:
• Batteries have been
removed while the
pump was on battery
source.
• External power has
failed or has been
disconnected and no
batteries are installed.
• Pump has been
powered on with
insufficient power
source.
Abbott GemStar Manual 430-600095-002 (Rev. 8/02)
Cause/
Remedy
1. Press [SILENCE] to
quiet the alarm for 15
minutes.
2. Press [STOP] to place
the pump in the STOP
mode.
3. Connect the pump to
AC mains adaptor,
rechargeable battery
pack, OR replace both
disposable batteries
with fresh AA batteries.
4. Press [START] to begin
delivery.
1. Connect the pump to
AC mains adaptor,
rechargeable battery
pack, OR replace both
disposable batteries
with fresh AA batteries.
2. Power-on the pump.
Review the program if
required.
3. Press [START] to begin
delivery.