Maintenance
Recommended Cleaning Solutions
The pump is not adversely affected by the recommended cleaning solutions in the table
below:
d
Note: NOT all cleaning solutions are disinfectants. Check product labeling.
Cleaning Solution
Super Edisonite®
Vesphene II se®
Manu-Klenz®
Formula C™
Household bleach
Cleaning the Pump Case
1
Power-off the pump.
2
Disconnect the pump from AC mains power.
3
Clean the exposed surfaces of the pump with a soft, lint-free cloth dampened with
the appropriate cleaning solution.
4
Wipe the solution from the pump surface with a water-moistened, lint-free cloth.
5
Use a dry, lint-free cloth to dry the pump after cleaning.
100
Manufacturer
Preparation
S. M. Edison
Per manufacturer's recommendation
Chemical Co.
Calgon Vestal
Per manufacturer's recommendation
Laboratories
Calgon Vestal
Per manufacturer's recommendation
Laboratories
Diversey Corp.
Per manufacturer's recommendation
Various
Per healthcare facility procedures; DO NOT exceed
one part bleach in ten parts water
Abbott GemStar Manual 430-600095-002 (Rev. 8/02)