Abbott GEMSTAR Operating Manual page 81

Table of Contents

Advertisement

Options Menu
Abbott GemStar
U = Unlock
T = Therapy Lock
R = Rate Change Lock
Function
Lock/Unlock Keypad
Power Pump On or Off
Start or Stop Infusion
Access Display, Help, Print, and Program
Review Functions
Deliver a Bolus
Silence an Alarm
Select a New Container
Purge
Change a Program
Review/Recall Speed Protocol
Assign or Delete Speed Protocol
Set Air Sensitivity, Distal Occlusion Level,
Sound Level, and Clock
Clear Shift Totals
Program and Deliver a Loading Dose
(Outside of Programming Mode)
Enter a New Program
Piggybacking
Rate Titration
Operation Test
Clear History
Select Another Therapy
74
Keypad Lock Level Functions
N = New Container Lock
F = Full Lock
= Function is Available
U T
R
N F
Comments
· · · · ·
Rate Change Lock is available only if the current
· · · · ·
program is Pain Management.
· · · · ·
· · · · ·
· · · · ·
· · · · ·
If programmed.
· · · ·
· · · ·
· · ·
Under Rate Change Lock, only the delivery rate or
bolus dose volume can be changed and must be within
the preset values.
· ·
Under Therapy Lock, the recalled protocol must be the
same type that was programmed when the pump was
· ·
locked.
· ·
· · · ·
· ·
· ·
Under Therapy Lock, the new program must be of the
same type that was programmed when the pump was
· ·
locked.
· ·
·
·
·
Abbott GemStar Manual 430-600095-002 (Rev. 8/02)