Changing a Pain Management Program

d
Note: You cannot change a program when the pump is in
With a
RATE CHANGE
within the set ranges.
Can be changed at any time
• Delivery rate
• Bolus
• Bolus lockout
d
Note: Although the delivery mode cannot be changed, if a Continuous + Bolus therapy is
programmed, the continuous rate or bolus amount can be changed to zero. This function may
be restricted by the minimum values set with a
To change a program:
1
From the
STOP
2
As each program parameter displays, make the desired changes, or press
[
] to accept the current setting.
YES/ENTER
3
Review the program.
4
Press [
START
Abbott GemStar Manual 430-600095-002 (Rev. 8/02)
Lock, the continuous delivery rate and bolus amount can be changed
• Container size
• Air sensitivity
• Dose limit amount
RATE CHANGE
mode, press [
] and select [
CHANGE
] to begin delivery.
Pain Management
or
FULL
NEW CONTAINER
Cannot be changed
• Delivery mode
• Unit of measure
• Concentration
• Dose limit type
Lock.
]
.
4
CHANGE PROGRAM
41
Lock.