Weight Dosed

Weight Dosed Programming Worksheet

GIVE:
Parameter
Therapy
Unit of Measure
Patient Weight
Concentration
(2 entries are required. Use
toggle)
Rate
VTBI
(Volume To Be Infused)
* Programmable Ranges for Rate:
0.1 - 8333 mg/kg/min
0.1 - 500000 mg/kg/hr
0.1 - 16666 mg/min
0.01 - 166666 mcg/kg/min
0.01 - 9999999 mcg/kg/hr
0.01 - 166666 mcg/min
0.1 - 1000 mL/hr
CAUTION:
When infusing short-half-life drugs (i.e. those dosed with units of
mcg/kg/min) at very low rates (0.8 mL/hr or less) the rate may be too slow
for that drug. Consider using a lower drug concentration with these
drugs so a pump rate above 0.8 mL/hr can be used.
36
Abbott GemStar
Selections and/or Values to be
Entered
Weight Dosed
q mcg/kg/min
q mcg/kg/hr
q mcg/min
q mg/kg/min
q mg/kg/hr
q mg/min
q mL/hr
(When programming in kg/min or kg/hr)
(For all units of measure EXCEPT mL/hr)
to
mg
in
mL
Abbott GemStar Manual 430-600095-002 (Rev. 8/02)
Programmable
Ranges
N/A
N/A
2.0 - 200.0 kg
0.1 - 9999000 mg
0.1 - 9999 mL
(see below *)
0.1 - 9999 mL

Advertisement