Tpn Programming Worksheet - Abbott GEMSTAR Operating Manual

Table of Contents

Advertisement

TPN Programming Worksheet

GIVE:
Parameters
Therapy
Delivery Mode
VTBI
(Volume To Be Infused)
Taper Up
Taper Down
Total Time to Infuse
KVO
(Keep Vein Open)
Container Size
(VTBI + KVO)
Air Sensitivity
* Maximum VTBI is 9600 mL minus any taper amount.
** Programmable ranges for taper down are 1 minute to 3 hours OR 2 times the remaining volume
divided by the continuous delivery rate, whichever is less.
Abbott GemStar Manual 430-600095-002 (Rev. 8/02)
Abbott GemStar
Selections and/or Values to be
Entered
TPN
q Continuous
q Continuous + Taper Down
q Continuous + Taper Up
q Continuous + Taper Up and Down
mL
q No
q Yes
hrs:min
q No
q Yes
hrs:min
hrs:min
q No
KVO Rate: 0.0
q Yes KVO Rate:
mL
q On
Alarms at approx. 0.5 mL of air
q 2 mL Alarms at approx. 2 mL of air
q Off
No alarm
Total Parenteral Nutrition (TPN)
Programmable Ranges
N/A
N/A
1 - 9600 mL*
1 min - 3 hrs
1 min - 3 hrs**
1 min - 24 hrs
1.0 - 5.0 mL/hr
9999 mL max
N/A