Troubleshooting
Alert Display Message
Y = Yes
N = No
END OF INFUSION
The pump delivers
KVO if selected.
FLASHING DISPLAY
Line four of the
display flashes.
PROGRAMMING
INCOMPLETE
START
START flashes on the
display. Other parts
of the display are not
affected.
96
Is the pump
Alarm
infusing?
Condition
A = Audible alarm
V = Visual alarm
N
A/V
The pump has
delivered the entire
programmed protocol.
N
A/V
A Non-programming
Callback has occurred
because the pump has
been waiting for more
than three minutes for
a keypress on a
non-programming
input or a confirmation
display.
N
A/V
Programming has not
been completed and no
keys have been pressed
within three minutes.
N
A/V
A Stop Mode Callback
has occurred because
the programmed pump
has remained in the
STOP mode for more
than three minutes.
Abbott GemStar Manual 430-600095-002 (Rev. 8/02)
Cause/
Remedy
If no KVO is selected:
1. Press [SILENCE] to
quiet the alarm for 2
minutes.
2. Press [STOP] to place
the pump in the STOP
mode and clear the
alarm.
If KVO is selected:
1. The pump sounds 5
beeps approximately
every 15 minutes.
2. Press [STOP] to place
the pump in STOP
mode and clear the
alarm.
Follow the instructions on
the display to clear the
alarm.
1. Press any key to clear
the alarm.
2. Complete the program
entries.
1. Press [SILENCE] to
quiet the alarm for 3
minutes.
2. Follow the instructions
on the display to clear
the alarm.