Troubleshooting
Alarm Display Message
Y = Yes
N = No
CHANGE BATTERIES
CHECK CASSETTE-X
CALL
1.800.XXX.XXXX
CODE: NN/MMM/TTT
90
Is the
pump
Alarm
Condition
infusing?
A = Audible alarm
V = Visual alarm
N
A/V
The pump has entered the
STOP mode due to
battery voltage below
acceptable levels.
N
A/V
The cassette is improperly
installed or not installed.
X indicates where to look
for the cause of the alarm:
A = Air Sensor
D = Distal
P = Proximal
N
A/V
The pump has detected a
condition that may require
service.
Abbott GemStar Manual 430-600095-002 (Rev. 8/02)
Cause/
Remedy
1. Replace both disposable
batteries with fresh AA
batteries and press
[
] to restart the
ON/OFF
pump; OR temporarily
connect the pump to
the AC mains adaptor
and the pump will
automatically restart.
Replace both batteries
as soon as possible.
2. After power-on, resume
the current program.
Refer to Accessing the
Programming Menu in
the "Basic Operation"
section for more
information.
1. Press [SILENCE] to
quiet the alarm for 1
minute.
2. Press [STOP] to place
the pump in the STOP
mode.
3. Reinstall the cassette.
4. Press [START] to begin
delivery.
1. Write down the codes
and telephone number.
2. Power-off the pump. If
the pump does not
power-off, disconnect
external power and
remove the batteries.
3. Call the telephone
number shown on the
display message.