Priming Pump Sets WITHOUT a Drip Chamber
WARNING: Failure to use the anti-siphon valve may result in unrestricted
flow.
1
Push down on the flow stop
is in the open (priming) position.
2
Hold the flexible fluid container with the administration
port pointed toward the ceiling and insert the piercing
pin.
3
Open the slide clamp.
4
Invert the cassette approximately 45°.
5
While holding the inverted cassette, squeeze or roll the
flexible fluid container to eliminate air from the
container, cassette, and tubing.
6
If necessary, tap the cassette to eliminate any remaining
air bubbles.
7
Return the cassette to the upright position. Continue
priming until fluid fills the tubing.
Abbott GemStar Manual 430-600095-002 (Rev. 8/02)
to ensure the cassette
System Setup
19