System Setup
8
Push down on the flow stop
is in the closed position.
9
Insert the cassette into the cassette pocket. Refer to
"Loading the Cassette" on page 26.
CAUTION: When priming is complete, ensure no fluid flows at the distal end of
10
Connect the pump set to the patient-access device.
Priming Pump Sets WITHOUT a Drip Chamber Using the PURGE Key
WARNING: Failure to use the anti-siphon valve may result in unrestricted
flow.
1
Push down on the flow stop
is in the closed position.
2
Hold the flexible fluid container with the administration
port pointed toward the ceiling and insert the piercing
pin.
3
Insert the cassette into the cassette pocket. Refer to "Loading the Cassette" on
page 26.
4
Press [
PURGE
5
While holding the flexible fluid container with the administration port pointing
toward the ceiling, press and hold [
d
Note: Purge rate is 250 mL/hr. The pump purges for up to 2 minutes each time you press
and hold [
PURGE
Continue using the PURGE key until fluid fills the tubing. Ensure all air is removed
from the container, cassette, and tubing.
CAUTION:
6
Connect the pump set to the patient-access device.
20
to ensure the cassette
the pump set. If flow is observed, DO NOT use the pump set.
to ensure the cassette
].
displays. Press [
PRIME THE SET?
PURGE
].
When priming is complete, ensure no fluid flows at the distal end of
the pump set. If flow is observed, DO NOT use the pump set.