Ict; L And H (Reference Range) Result Flags Display With No Result When Error Log Message 351 Or 352 Is Generated - Abbott AEROSET Troubleshooting Manual

Table of Contents

Advertisement

ICT™

L and H (Reference Range) Result Flags Display with No Result when Error Log
Message 351 or 352 Is Generated
®
AEROSET
System Troubleshooting Guide
94816-107—November 2004
PROPRIETARY INFORMATION
CONFIDENTIAL MATERIAL
Not for Customer Distribution
When the Error Log Message "351 - ICT Reference Solution
Cup Not Filling" or "352 - ICT Reference Solution Not
Aspirated" is generated, the ICT result is masked but the L and
H result flags display.
Resolution: There is no corrective action to mask the L and H
flag. Refer to Error Log Messages—Supplemental
Information in Section 10, Troubleshooting and Diagnostics of
the AEROSET System Operations Manual for initial recovery
from Error Log Messages 88, 351, and 352.